Command SJ-300F

Overview:

Key Features:

Graphical synoptic display with real-time system status. Full system alarming. Historical data trending, with quick export capability. Power outage protection including product recovery routines. Comprehensive programming of unlimited custom recipes. Semiautomatic and fully automatic modes of operation. Available optional features include automatic Batch Reports


21 CFR Part 11 Compliance CapabilitiesE-Signatures attached to all recipe changes. Secure Database alarms and events database. Proficy Historian for secured historical data archiving. Full security integration with the Microsoft Windows operating system environment


Specification

1.Command SJ-300F(T)

Manufacture Standards
A. Chamber
Name Model/Specifications/Material Number
01 Chamber AISI 304L /316L 1
2 Shelves AISI 304L /316L 1Set
500× 800mm 8+1
3 Chamber Door Acrylic or stainless door 1
4 Product  temperature sensors Pt-100 8
5 Temperature sensors(thermal fluid) Pt-100 2
6 Chamber material AISI 304L /316L 1
Performance parameters
1 Finish Ra≤0.4µm
2 Level 0.5mm/m
3 Shelves temperature Min <-60℃
4 Shelves temperature Max +65℃
5 Shelves temperature uniformity ±1°C(after balance)
6 Shelf cooling time (without load) +20 ~ -40°C ≤ 60 minutes
7 Shelves Distance 100mm(stopper 70mm)
8 Chamber type Rectangular
9 Insulation B1 Cotton insulation
10 Door Open direction Left or right
11 Sealing strips for doors Silicon rubber
B. Condenser (with coiled pipe)
SN Name Model/Specifications/Material Number
1 Condenser AISI 304 ,316 optional 1
2 Condenser coiled pipe Coil inside 1
3 Temperature sensors(condenser) Pt-100 1
4 Defrost water soaking defrosting
Performance parameters
5 Coiled temperature ≤-85℃
6 Condenser cooling time(without load) +20 ~ -40°C ≤ 30 minutes
D  Refrigeration system
SN Name Model/Specifications/Material Number
1 Compressor Frascold 6HP 2
2 Refrigerant R507A +R508B
3 Oil separator Emerson 1
5 Automatic protective device for pressure DANFOSS 2
6 Pressure meter REFCO 3
7 Expansion valve DANFOSS 3
E  Vacuum system
SN Name Model/Specifications/Material Number
1 Vacuum pump DRV90 1
2 Isolation valve of pump well Electric valve 1
3 Vacuum Sensor Posifa 2
4 CM sensor optional for Chamer
Performance parameters
1 Lowest vacuum pressure <=1Pa
2 Evacuation time From atm to 10pa≤30min
3 Leakage rate 5×10-3pa.m3/sec
G  Circulation system
SN Name Model/Specifications/Material
1 Circulation pump Low temperautre magnetic pump 1
2 Thermal fluid 5cs,USA Dowcorning Silicon oil (low viscosity)
3 Heater 2000w 2
4 Plate heat exchanger 2
Performance parameters
1 Heating speed 1°C/min
Electric control system
SN Name Model/Specifications/Material
1 Electric standard IEEE
2 PLC Simens S7-200 1
3 Breaker schneider
4 AC contactor schneider
8 Control type Manual and automatic
Software function
1 Freeze drying (switch the type between manual and automatic randomly)
2 Technique setting (recipe management)
3 Select Scada soft 21 CFR Part 11 Compliance CapabilitiesE-Signatures attached to all recipe changes.
1 Power requirement 25Kw (380V,-440v 50Hz/60hz, 3Phase)or 220-240v 60hz 3phase
2 Machine construction Wall mounted ,GMP type
3 Weight (appox) 2500kg
4 Cooling requirement water cooling(if pipe size is 1 inch,water pressure 2.5bar-3.5bar),Temperature <25C
Verification documents system The details of verification documents system
Operation manual 1
Wiring drawing 1
PQ,OQ,IQ Document 1

Picture for reference

image.png

image.png

image.png

image.png

image.png